Certain 802.11 network management messages have been determined to invoke wireless access point blacklisting security defenses when not required, which can necessitate intervention by hospital staff to reset the device and reestablish a network connection to the Wi-Fi access point. During this state, the Philips IntelliVue MX40 Version B.06.18 can either connect to an alternative access point within signal range for association to a central monitoring station, or it can remain in local monitoring mode until the device is reset by hospital staff. CVSS v3 base score: 6.5, CVSS vector string: AV:A/AC:L/PR:N/UI:N/S:U/C:N/I:N/A:H. Philips has released software update, Version B.06.18, to fix the improper cleanup on thrown exception vulnerability, and implement mitigations to reduce the risk…

CVSS v3.1
6.5 MEDIUM
EPSS
0.83%
probability of exploitation in 30 days
CISA KEV
Not listed
no confirmed exploitation reported
News coverage
0
sources referencing this CVE
EPSS Trend
Aug 16, 2026Sep 7, 2026
0.01pp
0.84% → 0.83% over 7 tracked changes
CVSS v3.1 Detail
Attack VectorADJACENT
Attack ComplexityLOW
Privileges RequiredNONE
User InteractionNONE
ScopeUNCHANGED
ConfidentialityNONE
IntegrityNONE
AvailabilityHIGH
CVSS:3.0/AV:A/AC:L/PR:N/UI:N/S:U/C:N/I:N/A:H
Connections
2 relationships
Timeline
disclosure → media coverage
Apr 30, 2018
Disclosure — published as a CVE record. · last revised by NVD Nov 21, 2024